    A method for source-depth estimation using a Hybrid Gamma Camera
EP23404
A method for source-depth estimation using a Hybrid Gamma Camera
Submitted on 25 Sep 2015

N.S.Dawood1, J.E. Lees1, S.L. Bugby1, L.K. Jambi1, M.S. Alqahtani1, W.R. McKnight1and A.C. Perkins2
1 Space Research Centre, University of Leicester, Leicester, LE1 7RH, UK 2 Radiological and Imaging Sciences, Medical Physics and Clinical Engineering, Medical School, University of Nottingham and Nottingham University Hospitals NHS Trust, NG7 2UH, UK
This poster was presented at BIR Annual Congress 2015
Poster Views: 877
View poster »
Poster Abstract
Objective: To estimate the depth of radioisotope distribution in relation to patient anatomy using a portable hybrid gamma camera which combines optical and gamma imaging. This additional information could be of particular help in optimising surgical procedures for resection of lymph nodes and tumours.

Method: The Hybrid Gamma Camera (HGC) was used to image phantoms filled with ⁹⁹ᵐTc to simulate radioactive uptake in nodes.
Gamma and optical images were recorded simultaneously and fused in a combined image. The distance between the simulated skin surface and the radiolabel in the nodes was determined from the images, and compared with the measured depth of the source below the surface.

Result: Good agreement was found between the calculated and measured source depths. Surface distance measurements were accurate to within 3% (at 200mm from the collimator face) and depth measurements within 10%. The phantom simulations have shown that this simple technique could provide a useful estimation of depth for radiolabelled distributions prior to clinical procedures. Suggestions for improving this technique will be discussed.

Conclusion: A method to calculate the depth from surface to source using the HGC has been described. There is a linear relationship between the calculated and measured distances from the camera to the source and/or the surface. This technique could allow surgeons to determine the depth and location of radiopharmaceutical uptake immediately prior to an operation without 3D SPECT imaging.


J E Lees, S L Bugby, A P Bark, DJ Bassford, PE Blackshaw, AC Perkins, A Hybrid Camera for locating gamma sources in the environment, JInst 8 (2013) P.10021 Report abuse »
Questions
Ask the author a question about this poster.
Ask a Question »

Creative Commons

Related Posters


Infarto agudo del miocardio en una paciente embarazada. Reporte de un caso.
Contreras Zárate M, Vázquez Rodríguez U, Jiménez Magaña F, Tetuán Medina E, Navarrete Ramírez H, Rodríguez R, Santoyo Saavedra A.

Potent transcriptional activation using CRISPRa with synthetic crRNA
Eldon T. Chou, Žaklina Strezoska, Elena Maksimova, Maren M. Gross, Melissa L. Kelley, and Anja van Brabant Smith

ADVISA PACEMAKER SCANNED AT 3 TESLA MRI SYSTEM: FIRST EXPERIENCE
Gabriele D’Ambrosio (1,2), Laura Anna Leo (2), Maria Luce Caputo (2), Francesco Fulvio Faletra (2), Maria Penco (1), Chiara Bucciarelli-Ducci (2,3)

Temporal trends and patterns in heart failure incidence: a population-based study of 4 million individuals
Nathalie Conrad, Andrew Judge, Jenny Tran, Hamid Mohseni, Deborah Hedgecott, Abel Perez Crespillo, Moira Allison, Harry Hemingway, John G Cleland, John J V McMurray, Kazem Rahimi.

Rede Nacional Cuidados Continuados Integrados - A Equipa Coordenadora Local Almada Seixal em articulação com as Equipas Referenciadoras do Hospital Garcia de Orta
Ana Henriques, Cidália Martins, Isabel Tavares, Manuela Pires, Micaela Abreu, Pedro Pacheco